Loading...
 
Toggle Health Problems and D

Big differences (~10 ng) in vitamin D test results, even when using the same type of tester – Aug 2015

Analytical Bias in the Measurement of Serum 25-Hydroxyvitamin D Concentrations Impairs Assessment of Vitamin D Status in Clinical and Research Settings

Lucinda J. Black , Denise Anderson, Michael W. Clarke, Anne-Louise Ponsonby, Robyn M. Lucas, Ausimmune Investigator Group
PLOS ONE 10.1371/journal.pone.0135478

VitaminDWiki Summary

50 blood samples were sent to four labs for vitamin D testing
Lab A (LC-MS/MS )vs Reference LC-MS/MS
Image

Lab B (LC-MS/MS) vs Reference LC-MS/MS
Image

Lab C (DiaSorin) vs Reference LC-MS/MS
Image

See also VitaminDWiki

What is the true vitamin D value? Mean ~ 48 nmol
1090 tests of THE SAME blood sample on one tester (figure 1)
Is the true value 48 nmol -1090 tests @ /is.gd/vitd1090
Thus can expect 5 ng of differences in test results from the SAME tester

Huge differences in % deficient (< 30ng) depending on the tester used

Image
Note: A, B, and C unable to detect/measure < 13 nmol (5 ng)

Measured serum 25-hydroxyvitamin D concentrations vary depending on the type of assay used and the specific laboratory undertaking the analysis, impairing the accurate assessment of vitamin D status. We investigated differences in serum 25-hydroxyvitamin D concentrations measured at three laboratories (laboratories A and B using an assay based on liquid chromatography-tandem mass spectrometry and laboratory C using a DiaSorin Liaison assay), against a laboratory using an assay based on liquid chromatography-tandem mass spectrometry that is certified to the standard reference method developed by the National Institute of Standards and Technology and Ghent University (referred to as the ‘certified laboratory’). Separate aliquots from the same original serum sample for a subset of 50 participants from the Ausimmune Study were analysed at the four laboratories. Bland-Altman plots were used to visually check agreement between each laboratory against the certified laboratory. Compared with the certified laboratory, serum 25-hydroxyvitamin D concentrations were on average 12.4 nmol/L higher at laboratory A (95% limits of agreement: -17.8,42.6); 12.8 nmol/L higher at laboratory B (95% limits of agreement: 0.8,24.8); and 10.6 nmol/L lower at laboratory C (95% limits of agreement: -48.4,27.1). The prevalence of vitamin D deficiency (defined here as 25-hydroxyvitamin D <50 nmol/L) was 24%, 16%, 12% and 41% at the certified laboratory, and laboratories A, B, and C, respectively. Our results demonstrate considerable differences in the measurement of 25-hydroxyvitamin D concentrations compared with a certified laboratory, even between laboratories using assays based on liquid chromatography-tandem mass spectrometry, which is often considered the gold-standard assay. To ensure accurate and reliable measurement of serum 25-hydroxyvitamin D concentrations, all laboratories should use an accuracy-based quality assurance system and, ideally, comply with international standardisation efforts.
 Download the PDF from VitaminDWiki

Attached files

ID Name Comment Uploaded Size Downloads
5796 Less than 30 ng.jpg admin 13 Aug, 2015 27.80 Kb 1857
5795 Lab C.jpg admin 13 Aug, 2015 23.71 Kb 1171
5794 Lab B.jpg admin 13 Aug, 2015 23.53 Kb 1322
5793 Lab A.jpg admin 13 Aug, 2015 26.76 Kb 1516
5792 Testing problems.pdf admin 13 Aug, 2015 685.69 Kb 952